LEPU / CREATIVE UP-7000

Multifunctional patient monitor, uninterrupted, 12.1 inch color LCD display.

Health care products / Patient Monitors


Multifunctional medical instrument for measuring and monitoring the physiological vital signs of adult and child patients.

- External (non-invasive) blood pressure measurement (NIBP).
- Manual, automatic and statistical measurement mode.
- Automatic measurement at intervals of 1-240 minutes (adjustable).
- Automatic measurement interval deviation: maximum 10 seconds.
- It can be adjusted to measure the blood pressure of adults, children and newborns.
- Blood pressure measurement with a cuff that can be pulled on the arm (oscillometric measurement).
- Cuff inflation time: maximum 30 seconds (for an adult).
- Pressure limit during heart contraction (systole): 25-265mmHg (adult and child), 25-135mmHg (neonate).
- Pressure limit when the heart relaxes (diastolic): 12-220mmHg (adult and child), 12-110mmHg (newborn).
- Mean arterial (MAP) pressure range: 18-245mmHg (adult and child), 18-120mmHg (neonate).
- Overpressure protection limit: 300mmHg (adult and child), 140mmHg (newborn).
- Blood pressure measurement margin of error: +/-10%.
- Maximum deviation from standard: +/-8mmHg.
- Two-point body temperature measurement.
- Body temperature measurement range: 15°C-45°C.
- Body temperature measurement accuracy: +/- 0.2°C.
- Respiration monitoring (RESP) monitoring range: 0rpm-120rpm.
- Respiration monitoring (RESP) alarm range: 0bpm-100bpm.
- Breath monitoring accuracy and alarm value limit: +/-10%.
- Seven-channel heart rhythm curve (ECG) monitoring.
- Heart rate measurement range: +/-0.4mVp - +/-3mVp.
- Heart rate measurement range: 20bpm-250bpm.
- Heart rate measurement accuracy: +/-2bpm or +/-1% (whichever is greater).
- Heart rate alarm delay time: maximum 12 seconds.
- Sensitivity setting: x1/2, x1, x2, x4.
- Sensitivity tolerance: maximum 10%.
- ECG noise level: maximum 30uVp-p.
- ECG input loop current: maximum 0.1uA.
- ECG input resistance: greater than 5Mohm.
- Irregular heartbeat (arrythmia) and ST segment analysis.
- Measurement of blood oxygen saturation (oxygen saturation, SpO2).
- Transducer: two wavelength LEDs.
- Blood oxygen saturation measurement range: 35%-100%.
- Blood oxygen saturation measurement accuracy: 0.4%-5%.
- Pulse waveform measurement range: 30bpm - 240bpm.
- Pulse waveform measurement accuracy: +/-2bpm or 2% (whichever is greater).
- Optional blood carbon dioxide pressure measurement and monitoring.
- Storage and viewing of 800 measured data.
- Storage and display of 24 hours of heart rhythm curves (ECG).
- 480 hours of graphical and tabular statements for all measured parameters.
- 12.1-inch color TFT display.
- 5-channel waveform and display of all parameters.
- Protection against defibrillator and electrocautery interference.
- Alarm with sound and light signal.
- Ethernet computer network connection to LEPU CMS central monitoring system (optional).
- A rechargeable battery ensures uninterrupted power supply (not included).
- Built-in thermal paper printer for printing graphs and text information (not included).
- Supply voltage: 100-250VAC, 50/60Hz.
- Power requirement: maximum 80VA
- CE certification.